Argos Technologies Flexifuge
Mini Centrifuge — Operator’s Manual
English
Troubleshooting
Error Cause Solution
Unit does not start Power cable not
connected
Check that power cable
is rmly plugged into both
wall outlet and device
Power supply Check that there is power
to the wall outlet
Lid not closed properly Press rmly to close lid
Power switch Check that power switch is
at the ON position
Excessive noise
or vibration
Rotor is not loaded
symmetrically
Stop the unit and load
the tubes symmetrically.
Ensure that the tubes on
the opposite sides are
equally lled.
Rotor is slanted Remove the rotor and
reinsert straight and rmly
downwards
Loose locking screw Check to see if locking
screw has been inserted
properly and that it has not
been over-tightened
Unit base is not level Check that unit on a at
surface
If suggested solutions are unsuccessful,
please contact Argos Technologies at 847-622-0456
Warranty
• e full warranty for this device is 12 months from the date of sale.
• During this period of time the device will be repaired or exchanged without charge.
• Warranty is null and void on devices with damaged manufacturer’s number; damaged seal;
damage due to improper use; storage, or transport of device; damage due to service by
unauthorized persons.